Abstract
Noise performance of a phase-locked loop (PLL) based frequency modulation Kelvin force microscope (FM-KFM) is assessed. Noise propagation is modeled step by step throughout the setup using both exact closed loop noise gains and an approximation known as “noise gain” from operational amplifier (OpAmp) design that offers the advantage of decoupling the noise performance study from considerations of stability and ideal loop response. The bandwidth can be chosen depending on how much noise is acceptable and it is shown that stability is not an issue up to a limit that will be discussed. With thermal and detector noise as the only sources, both approaches yield PLL frequency noise expressions equal to the theoretical value for self-oscillating circuits and in agreement with measurement, demonstrating that the PLL components neither modify nor contribute noise. Kelvin output noise is then investigated by modeling the surrounding bias feedback loop. A design rule is proposed that allows choosing the AC modulation frequency for optimized sharing of the PLL bandwidth between Kelvin and topography loops. A crossover criterion determines as a function of bandwidth, temperature and probe parameters whether thermal or detector noise is the dominating noise source. Probe merit factors for both cases are then established, suggesting how to tackle noise performance by probe design. Typical merit factors of common probe types are compared. This comprehensive study is an encouraging step toward a more integral performance assessment and a remedy against focusing on single aspects and optimizing around randomly chosen key values.
Introduction
Surface potential imaging in combination with atomic force microscopy in ultrahigh vacuum is based on the measurement of electrostatic forces in amplitude modulation Kelvin force microscopy (AM-KFM) [1] or the measurement of the electrostatic force gradient in FM-KFM [2], in analogy with the FM mode used in noncontact atomic force microscopy (nc-AFM) [3]. The FM-KFM mode is often favored either because when a higher derivative of the probe–sample capacity is used, it is expected to be more sensitive to the very extremity of the tip [4], or because the use of probes with an increased fundamental resonance frequency makes the use of higher harmonics for simultaneous surface potential imaging inaccessible to the bandwidth of the deflection detector.
Previous studies of noise propagation often retrieve the general expression of frequency noise of a thermally excited harmonic oscillator and are not specific to a PLL based setup, and furthermore, do not extend to the noise in the KFM signal. The pioneer work on nc-AFM, [3] already mentions frequency noise for the first time in the context of nc-AFM, but takes into account only thermal probe excitation noise. Fukuma et al. [5] performed a detailed study on optimizing the probe deflection sensor and compare the measured noise power spectral density (PSD) at the PLL frequency output to the theoretical values derived from both thermal probe excitation and deflection sensor noise. Kobayashi et al. [6] focus on noise propagation in low quality factor (low-Q) environments for the application in liquids. Polesel-Maris et al. [7] studied the noise propagation in both amplitude and phase feedback loops of a nc-AFM as a function of the feedback controller settings, and showed that at a weak probe–surface interaction, the feedback loops can be considered independently whereas at a strong interaction, they become coupled. In our work on the dynamic behavior of AM-KFM [8], we studied the noise propagation from sensor displacement noise to the Kelvin voltage output. Giessibl et al. [9] compared qPlus and length-extension resonator (LER) sensors with respect to four noise sources: thermal excitation, sensor displacement noise, oscillator noise and thermal drift noise. The impact of all noise sources on frequency noise was discussed. Finally, Lubbe et al [10] numerically modeled noise propagation from sensor displacement noise to frequency noise of a PLL based nc-AFM depending on filter settings.
In this work, the noise propagation of a PLL based FM-KFM is studied by measuring and analytically modeling noise at different stages of the setup starting from the beam deflection signal, via the phase detector and the PLL outputs up to the Kelvin output voltage. The concept of noise gain allows for decoupling noise performance from the optimization of bandwidth and stability. It is commonly used in designing operational amplifier circuits. The noise PSD is modeled as if the bandwidth was unlimited and later, the bandwidth is chosen as a function of the acceptable signal fluctuation. This approach is appropriate because (1) increasing the closed loop bandwidth of a stable feedback loop above a certain frequency does not alter the noise PSD shape at the onset up to that frequency, and (2) stability and bandwidth are in many cases, including the described setup, not the bottleneck, i.e., constant gain can easily be achieved up to a frequency above the one at which the total output noise exceeds an acceptable value. The modeled noise PSD is in agreement with the measured one, showing that no significant noise contribution is added by the PLL. Since in FM-KFM the frequency shift signal is shared by both distance and potential control loops, a design rule for choosing the AC modulation frequency is proposed that ensures making best use of the available PLL bandwidth with negligible crosstalk between the loops and that yields equal bandwidth for both loops. The Kelvin output noise reduces to a compact analytic expression in terms of probe merit factors and a criterion for the transition between dominating detector and thermal excitation noise is derived. Noise optimization can then be approached via probe design after identifying the bottlenecks and addressing the respective parameters. The work is an approach toward a more integral view of KFM performance. A limit to optimization is the complicated interdependence of probe and detector parameters that for a practical implementation prevent reaching the ultimate theoretical limit imposed by the uncertainty principle.
Gain and noise gain
For studying the noise propagation across the control loops, the concept of noise gain from OpAmp circuits is adopted. Figure 1 shows an OpAmp in a typical configuration and its decomposition into forward and feedback gain, adder and noise source. Later on, each of the KFM control loops will be represented by a similar equivalent circuit. Generally, the feedback gain F corresponds to a PI (proportional, integral) controller.
The output signal Out is written as function of the input signal In, the noise An and the gains:
In this case the signal gain is equivalent to the noise gain
Depending on where the noise generator is inserted in the loop, the gains for signal and noise can be different as will be shown later. The approximation, although valid only in the operating bandwidth below the closed loop cutoff frequency, is widely accepted as the noise gain. The reason will be explained later.
The PLL controller
Figure 2 shows the setup of the PLL and the attribution of its components to the blocks A and F similar to Figure 1. The input is the resonance frequency variation Δf of the tip, which is subject to external influence (van-der-Waals or electrostatic tip–sample interaction), and which is to be tracked by a numerically controlled oscillator (NCO) that drives the piezo dither. To match the oscillator to the resonance frequency of the tip, the deflection of the tip is detected, and the phase shift with respect to the drive signal is determined by a lock-in amplifier. The phase shift is compared to a setpoint, and the error signal is amplified by a PI controller that controls the NCO with the objective of keeping the drive frequency matched to the resonance frequency. A perturbation can be injected to an input of a signal adder (as indicated) to study the loop response, or by modulating the resonance frequency of the probe, e.g., by exposing it to an electric field, which shall both yield the same closed loop response.
Phase detector gain - phase as function of frequency shift
We shall study the phase difference between a passive oscillator and a frequency modulated drive signal. If a resonator described by a quality factor Q and a resonance frequency f0 is excited by a frequency modulated drive force with an excursion fexc and a modulation frequency fpert:
The phase shift is, without any assumptions about frequency excursion, width of the resonance peak, or modulation frequency exactly:
This can be derived heuristically by knowing that the phase is the integral over frequency difference in the regime of high modulation frequency fpert, but that the phase shift is capped by the extrema of the arctan function in the regime of steady excitation since one oscillator is passive. The same result had been found by Portes et al. [11] by solving the differential oscillation equation. This general equation yields the approximations for particular cases below that are so frequently found in the literature. It is noteworthy that the phase is generally complex, i.e., the phase difference is itself dephased with respect to the frequency modulation at fpert.
For this result, it is irrelevant whether the frequency difference is the result of applying a perturbation at the entrance of the NCO or of detuning the cantilever frequency. Since our digital AFM controller does not provide the option of modulating the excitation frequency, we will study the PLL response by perturbing the resonance frequency of the tip by applying a voltage between tip and sample. The first task is to determine the frequency shift induced as a function of the voltage and the fixed tip–sample distance of some tens of nanometers for the static case fpert = 0.
Figure 3 shows the frequency shift Δf as a function of the voltage, measured by acquiring a resonance curve per voltage value (black squares and black solid line parabola fit). It also shows the static phase shift under excitation at constant frequency f0 (red squares), which is then shifted from the actual resonance by Δf due to the influence of the electric field. Then, Equation 4 reduces to
The red solid line is an arctan fit according to Equation 5. Note that the two branches of arctan functions do not intersect exactly at zero phase. This occurs if the resonance frequency of the tip drifts above the excitation frequency during the measurement. Consequently, the possible phase excursion may be higher than 90°.
Next, the forward response of the PLL, APLL, is studied dynamically. This experiment has to be performed by applying a frequency modulation indirectly since the integrated lock-in module does not allow transfer function measurements by introducing a Δf perturbation. For doing so, the tip is excited at constant frequency f0 = 61.835 kHz. Then, the resonance frequency is modulated by applying a bias containing both a DC and a smaller AC component of 0.8 V and 0.2 V respectively, and the phase detector output is recorded as function of modulation frequency of a small AC bias. We set the DC and the AC voltage components to aim at a frequency excursion of around fexc ≈ 2 Hz as indicated by the arrow in Figure 3. The result is the spectrum shown in Figure 4 by black squares, giving the phase shift as a function of the modulation frequency.
It is fitted with Equation 4 multiplied by a lowpass function of the phase detector output filtering, a 2nd order Butterworth with fc,LP = 2500 Hz cutoff frequency.
The best fit is obtained for an excursion of fexc = 1.9 Hz and the previously found values for f0 = 61.835 kHz and Q = 22800 (see Experimental section).
For the following, a linear conversion gain of the phase detector must be defined in terms of phase divided by frequency excursion, as function of modulation frequency. Before we can divide Equation 4 by fexc, it is compulsory to approach the arctan function by its argument for small excursion, fexc < f0/(2Q), since the definition of a gain implies a linear dependence. Then, Equation 4 simplifies and dividing by the excursion yields:
The approximation of the arctan function by its argument for small excursion is at the very limit of validity here because f0/(2Q) = 1.35 Hz and fexc = 1.9 Hz. However when the phase detector is ulteriorly used in the closed PLL loop within its tracking bandwidth, the error is negligible: The closed loop gain is near unity in this range, meaning that the oscillator follows the (detuned) resonance frequency, and the frequency error remains at a fraction of the frequency excursion. The forward gain Equation 7 will be used to model both the closed loop PLL response and the shape of frequency noise PSD. It is also shown on Figure 4 as red curve. At the same time, the reciprocal value of the feedback gain FPLL is displayed. The feedback circuit is a PI (proportional, integral) amplifier with the following response FPLL:
The controller software automatically sets the time constant of the phase locked loop PI amplifier equal to the time constant of the phase detector lowpass function, τPLL = 2Q/2πf0, and the P gain such that the crossing with the forward gain occurs at the chosen PLL bandwidth of 1 kHz.
The 2.5 kHz lowpass HLP of the phase detector output is also a consequence of the choice of 1 kHz PLL bandwidth and automatically set by the controller. In this case, the feedback parameters were PPLL = 8.73 Hz/° and τPLL = 112 ms.
PLL noise
First, a noise spectrum is measured at the output of the photodetector, without probe excitation. It is shown in Figure 5.
The deflection noise spectrum Dn(f) contains a component due to thermal probe excitation plus a component due to detector output noise. The latter can be assumed to be constant over the relatively small frequency interval of the spectrum. The respective power spectral densities (PSD) in units of m/ are uncorrelated and hence add in quadrature. The noise PSD Dn(f) is therefore described by a quadrature sum of detector noise zn,S and a Lorentzian component of the same Q and f0 as the resonance curve previously determined by the microscope controller:
A curve fit with Equation 9 yields k = 1.2 N/m, Q = 22800, f0 = 61835 Hz and zn,S = 2·10−13 m/.
The decomposition is also indicated in Figure 5. The optical beam deflection conversion gain leading to the scale was calibrated by using the method of reduced frequency shift [12] and was 0.15 nm/mV. The principle of this method is to maintain a constant reduced frequency shift by varying simultaneously the excitation amplitude and the frequency shift setpoint of the noncontact mode following a certain algorithm. Then, the lower turning point of the tip remains equidistant from the sample surface, and the motion of the z-piezo represents the shift of oscillation amplitude as response to varying excitation amplitude.
Next, the noise propagation throughout PLL and Kelvin loop are studied. In order to be able to model the noise by the approach of noise gains as in Figure 1, it is necessary to present it by a noise source inserted between blocks APLL and FPLL. We shall now calculate how the displacement noise at the photodetector output transforms into phase noise at the phase detector output, which is represented by the phase noise generator of Figure 2. Figure 6 shows the vector diagram in the complex plane of a signal D0cos(2πf0t), representing the tip deflection, plus a spurious small signal acos(2πf1t) as a representation of the deflection detector noise. If demodulated by a lock-in amplifier at f0, in the reference system rotating at f0, the D0 vector is stationary and a is rotating around the end of D0 at f1 − f0.
The imaginary projection is then
The phase is for a << D0
A small signal vector a at f1 = f0 + fpert causes a phase oscillation at fpert. If a was rotating at f1 = f0 − fpert, it would also cause a phase oscillation at f but with opposite sign, and hence two vectors a at opposite difference frequencies would add arithmetically and cancel. Regarding the phase noise at a frequency fpert, the spurious superimposed oscillations are replaced by the respective noise densities at frequencies Dn(f0 ± fpert)[V/]. Since the two noise components are uncorrelated, the densities add in quadrature:
The factor 1/2 applies because half of the power spectral density (PSD) is projected onto each real and imaginary axis in the complex plane. In analogy with Equation 11, the phase noise PSD , becomes:
Due to the symmetry of the Lorentzian with its high quality factor, it is sufficient to use one branch of the Lorentzian, and expressed in degrees we obtain:
This expression gives the phase noise with the use of a lock-in amplifier. It had been derived in a similar way by Rast et al. [13]. It may not be valid for other phase comparators, e.g., edge triggered ones. It is basically a translation by f0 of the deflection noise PSD. The translation of the Lorentzian component of the deflection noise yields a first order lowpass with respect to fpert with a cutoff frequency f0/(2Q), whereas the constant detector shot noise zn,S is invariant under translation, yielding for the total phase noise the quadrature sum:
Figure 7 shows the measured phase noise at the phase detector output in open PLL, and in green the fit according to Equation 15. The parameters zn,S, f0, Q and k were kept identical to the ones of the curve fit of Figure 5, whereas the oscillation amplitude had to be adjusted to D0 = 0.85 nm. The red and blue lines show the fit decomposed into thermal excitation and detector noise components, respectively.
PLL closed loop gain
With the known transfer functions APLL from Equation 7 and F from Equation 8, the closed loop response of the PLL can be computed. For the equivalent circuit of Figure 2, we find a signal gain
which is plotted along with the measured response in Figure 8. The computation is performed on complex transfer functions and the plot only shows the modulus of the result. Note that in Equation 16, APLL and FPLL are complex and only in the end of the computation, the module is calculated.
PLL closed loop noise
With the known forward and feedback gains of the PLL loop, the closed loop noise output spectrum fn of the PLL is modeled. Since the noise source of Figure 2 is located differently between blocks A and F than that of Figure 1, the noise gain is different from the signal gain in contrast to the operational amplifier example, and writes
The frequency noise PSD fn at the output of the PLL is
The PLL phase detector output noise PSD modeled by Equation 15 is used as input. The calculation of the noise gain is also performed on complex gains.
Figure 9 shows the PLL closed loop PSD of noise fn, (black), up to 500 Hz, the limit of the integrated spectral analyzer, and the numerically computed noise PSD (green) obtained from phase noise and gains APLL and FPLL. Furthermore, the noise PSD decomposed into thermal noise (red) and sensor noise (blue) is computed according to the approximation of Equation 17 (see Equation 20 below).
Regarding the approximation of Equation 17, we note that it is valid in the range in which the closed loop gain is unity and |APLLFPLL| >> 1. Hence, the approximation does not predict the roll-off of the noise PSD beyond the closed loop bandwidth. Instead, it predicts infinite rise of the noise following the blue line of Figure 9. To use the approximation 1/APLL, denoted “noise gain”, rather than the exact computation is the idea of the noise gain formalism and is justified by the following arguments:
- The roll-off of the noise PSD cannot be exploited anyway: if the loop is used beyond its cutoff frequency, it attenuates the signal as much as the noise. The image acquisition circuitry that samples data into pixels has an anti-aliasing low-pass filter with a cut-off frequency of half the sampling rate. There is no interest of using a loop bandwidth below that cut-off frequency since the response of the loop would then smooth the image at the same rate as it would smooth out noise.
- If the loop is inserted into a surrounding loop, then the closed loop gain of the former becomes the forward gain of the latter; consequently, the roll-off has two effects that compensate each other: first, it cuts off the noise PSD, but second, since the reciprocal of the inner loop closed loop gain becomes itself the (approximate) noise gain of the surrounding loop, it would amplify the noise PSD by as much as it had been attenuated before. Therefore, it is convenient to neglect the cutoff in noise propagation.
- Last, it is noteworthy that the closed loop cutoff frequency has no influence on the noise PSD at the onset below that cutoff frequency, e.g., the noise PSD from zero to 300 Hz is the same irrespective of whether the closed loop cutoff frequency is 500 Hz or 1 kHz. Therefore, it is convenient to first calculate the noise PSD as if the closed loop bandwidth was infinite, to determine over which frequency range the noise PSD can be integrated without exceeding an acceptable total signal fluctuation, and to limit bandwidth and sampling rate a posteriori. It will be discussed later to what extent the approach is feasible and whether stability issues can become the bottleneck.
Therefore among engineers the noise gain formalism is widely used but to our knowledge has not yet been applied to noise propagation in scanning probe microscopy. The PLL output noise PSD is now obtained using the noise gain formalism:
Regarding the PLL forward gain APLL, Equation 7, it is noteworthy that the open loop gain of the phase as function of frequency excursion has the same frequency dependence as the thermal contribution of the phase noise, second term of Equation 15, i.e., a first order lowpass with cutoff frequency f0/(2Q). The quotient Equation 19 yields, when inserting the phase noise PSD from Equation 15 and PLL forward gain APLL from Equation 7:
Hence, the thermal part of the frequency noise, is exactly constant over a range from zero to infinite frequency, which is the third term. It was derived by theorists and resumed by Giessibl and Kobayashi that a thermally excited harmonic oscillator is expected to have constant frequency noise PSD [5,14]. Here, we have provided the comprehensive step-by-step evidence for an experimental PLL setup with a driven passive resonator, yielding the same result. Controversial debate about the frequency noise of comparable PLL setups is still ongoing [15,16].
The sensor noise contribution is split into two contributions, the first term without frequency dependence, the second rising with fpert above f0/(2Q), meaning above 1.35 Hz in our case, below which it has a plateau. We state that Equation 20 is identical to Equation 18 of reference [6], up to a factor of 2 in front of the sensor noise contributions. The first frequency independent term of our sensor noise has been referred to as “oscillator noise” by Kobayashi which was later also adopted by Giessibl [9]. Following our approach, it is arising merely from propagation of sensor noise throughout the PLL. This frequency-independent component of sensor noise is only found in modeling if the PLL forward gain, Equation 7, is derived exactly with the f0/(2Q) corner frequency, rather than an approximate 1/fpert behavior. It is generally negligible in high-Q environments as in our setup. The good agreement between computed noise and experiment shows that here the frequency noise can be attributed solely to thermal excitation and sensor noise.
The Kelvin loop
Figure 10 shows the setup of the Kelvin loop. The closed PLL loop of the previous section now presents a small part of the forward gain. The AK block further contains a lock-in amplifier working at a frequency lower than the bandwidth of the PLL that modulates the gap voltage. It superposes an AC signal Vmod at a frequency within the operating range of the PLL and detects the resonance frequency modulation of the tip at this frequency. The output of block AK is the demodulated frequency shift Δfdemod, while the input is the VK component of the tip bias. The tip voltage superposition has two purposes: first to extract the polarity information of the gap voltage mismatch, and second to share the PLL bandwidth between Kelvin and distance controller: van-der-Waals and electrostatic interaction both shift the resonance frequency. A modulation of Δf at a frequency within the bandwidth of the distance controller would cause the distance controller to retract the tip periodically. Therefore, by modulation and demodulation, the electrostatically induced tip frequency modulation is translated in a range above the cutoff frequency of the distance controller, but below the cutoff frequency of the PLL.
The contact potential difference (CPD) between tip and sample is indicated by a voltage applied to the sample. It may be due to a work function difference between the sample and the tip or due to a sample to which a bias is applied. The objective is to cancel the CPD by applying a Kelvin voltage to the tip such that CPD − VK = 0.
Open loop forward gain
The CPD, the Kelvin voltage, VK, and the AC voltage, Vmod, cause an electrostatic field gradient that alters the resonance frequency:
The assumption of a constant d2C/dz2 is an approximation for oscillation amplitudes smaller than the mean tip–sample distance. However, the voltage dependence is valid even if this condition is not exactly met. By expanding the square bracket of Equation 21, one gets
The term of interest is the mixed term at fmod since it contains amplitude and sign of the CPD. It is detected by demodulating at fmod. The PLL response HPLL(fmod) applies before the demodulation. The static forward gain AK,DC is:
The static forward gain is determined with engaged distance control loop while using a setpoint of Δf = −5 Hz with Vmod = 300 mV, and fmod = 200 Hz. The demodulated error signal is then measured as a function of ΔVK, shown in Figure 11. The gain is 25 Hz/V.
Next, the forward gain is studied dynamically. Therefore, the DC voltage mismatch CPD − VK is replaced by an AC voltage Vpertcos(2πfpertt) and Equation 22 becomes
The mixed term will transform into two satellites at fmod ± fpert and Equation 23 becomes
If the PLL response is flat and unity around fmod ± fpert, above expression is equal to the static gain, multiplied by the output filtering of the Kelvin lock-in amplifier Hlockin. The validity of Equation 25 requires that the distance control loop does not interfere with the Kelvin control loop. First, it must not modify, by tip–surface interaction, the PLL response, e.g., by modifying Q via dissipation; second, it must not respond periodically to the frequency modulations caused by the Kelvin loop. This means that fmod − fpert must be above the cutoff frequency of the distance control loop. The ranges of PLL bandwidth occupied by distance and Kelvin loop are indicated by the arrows in Figure 8. In a range of fpert where the PLL closed loop has unity gain for fmod ± fpert, Equation 25 can be approximated as
Noise projection behind the Kelvin lock-in amplifier
Concerning the noise, if an equivalent control loop circuit in the sense of Figure 1 is to be applied, it is required to express the noise PSD at the interface of block A of Figure 10, meaning in the X output of the Kelvin lock-in amplifier. Hence the propagation of the noise PSD of the PLL output to the output of the Kelvin lock-in is now calculated. The projection of the PLL noise to the demodulated X output is the average between the satellites at fmod ± fpert
where fn is given by Equation 20.
Kelvin closed loop gain and noise
The loop is closed with a feedback gain FK(fpert) of a PI controller described by Equation 28:
The feedback parameters are set to PK = 1 mV/Hz and τK = 200 µs. To understand the choice of the parameters and their effect on the closed loop response, a schematics depicting both forward response |AK| and reciprocal of the feedback response, 1/|FK|, is shown in Figure 12.
The main point is that the 1/FK responsecurve crosses, with its slope, the AK responsecurve at a frequency where it is essentially constant. Many combinations of PK and τK are possible that yield the same closed loop cut-off frequency because only the frequency of crossing matters, but not the height of the plateau of the |1/FK| function. In the P–I representation, the IK component would need to be set to a specific value while the PK could be varied in a wide range. With the known open loop forward gain and output noise PSD (Δf)n of the PLL, it is possible to calculate the closed Kelvin loop signal and noise gain according to Figure 10 to compare them to the measured spectra:
The results are shown in Figure 13 and Figure 14 respectively. The fits have been obtained by using as output filtering of the Kelvin lock-in, Hlockin(fpert), a second order function with cutoff at 60 Hz. This cannot be set manually in our case and is thought to be directly coupled to fmod = 200 Hz.
The observation that the closed loop Kelvin response measured with engaged distance control is in agreement with the modeling based on the PLL response determined with retracted tip, supports the assumption that the gain and distance control loops with a setpoint of Δf = −5 Hz do not interfere with the Kelvin control loop by modifying the forward gain of the PLL. This situation corresponds to a weak surface interaction in the sense of [7].
The green curve of Figure 14 shows the numerically computed noise according to the exact expression of Equation 30 and Equation 31 and the demodulated noise of Equation 27. The red and the blue curves are the decomposed thermal and sensor noise calculated from the approximate noise gain of Equation 31 and the demodulated noise of Equation 27. Both computations do not reflect the little harmonic overshoot of the spectrum at around 25 Hz. Again, the approximation by using noise gain is accurate only up to the roll-off of the closed loop response. The sensor noise component of VK,n is relatively constant in contrast to the rising sensor noise at the PLL output because it is the average between the satellites at fmod ± fpert.
Discussion
Up to here, a typical laboratory setup has been treated in order to validate numerical and analytical treatment of noise propagation. Here, the cutoff frequency of the Kelvin loop had been set to 30 Hz by the choice of the feedback parameters as shown in Figure 12. With a Kelvin noise PSD of around 4 mV/, the total noise is expected to be around 22 mV. The bandwidth is an arbitrary choice and is limited by the acceptable noise level. The PLL bandwidth could indeed be set to a value in the kHz range, allowing to increase the AC modulation frequency and bandwidth of both distance and Kelvin control loops. In the following, the constraints with respect to a maximum bandwidth to noise performance shall be addressed.
Choice of fmod with respect to bandwidth BW
A design rule for the choice of the different frequencies is given in Figure 15: the black (solid) curve schematically represents the gain of the PLL controller. The red (dashed) curve is the gain of the distance controller. The green (dotted) curve is the range in terms of PLL frequency occupied by the Kelvin loop, consisting of two satellites of the Kelvin response around the AC modulation frequency. It is reasonable to plan the bandwidth of the distance control loop to be equal to the one of the Kelvin controller, fc,AFM = fc,KFM = BW, since usually both images are sampled at the same rate because it is a one pass technique and the Kelvin image is typically acquired with the same resolution as the topography image. If the modulation frequency is chosen to be fmod ≈ 4fc,AFM, then the Kelvin loop is using the PLL in a frequency range up to fmod + fc,KFM = 5 BW, which should be at a value such that the total noise remains acceptable (see section “Kelvin voltage noise PSD”), and on the other hand, the overlap and hence crosstalk between topography and KFM image is small since the roll-off of the distance controller at 1 BW and of the lower PLL frequency satellite of the Kelvin controller at fmod − fc,KFM = 3 BW are 2 BW apart. The cutoff frequency of the PLL, the AC frequency and the bandwidth of the Kelvin loop can be set to much higher values as discussed in the section “Absolute frequency limits irrespective of noise”. However the effective noise PSD of the PLL is composed of thermal and detector noise, shown in Figure 5, Figure 7 and Figure 9 and by Equation 15 and Equation 20. Care has to be taken that total noise, i.e., the integral of the noise PSD over the operating range, remains acceptable.
The 2 BW gap between the roll-off frequencies, together with the finding that both closed loop responses are second order systems, ensures that a CPD represented by Vpert varying at fc = BW causes a response of the distance controller at 3fc at −24 dB below its response to a static CPD. According to Equation 24, an AC CPD represented by Vpert, as well as the AC voltage Vmod, also both introduce a static term. The crosstalk onto the distance controller of Vpert oscillating at fc introduces an oscillation of z at 3fc:
The first two factors are
and
which is highly non-linear and dependent on the Δf setpoint. It is sufficient to know the product. This can be determined from the static term:
It is appropriate to choose a Δf setpoint such that the tip does not retract considerably in response to the applied value of Vmod. Equation 24 contains terms that cause a static tip retraction, and a dynamic tip movement at 3fc and 5fc due to mixed terms, and at 2fc and 8fc due to squared terms. If amplitudes are equal, VmodVpert = = , then the dynamic tip retraction at 3fc is −24 dB below the sum of the constant terms, or 6%. The constant tip retraction can be thought to be less troublesome because it introduces only an offset in the topography image while the retraction from varying surface potential introduces a real artifact. Nevertheless, it is favorable to minimize the tip–sample distance since it deteriorates the lateral resolution. Setpoint Δf and Vmod should be chosen such that the topography feedback is still dominated by van-der-Waals interaction. However, the tip–sample separation cannot be made infinitely small by hardening the topography feedback because of the snap-to-contact phenomenon. The ultimate limit is discussed below, and constraints between tip–sample separation, oscillation amplitude, and Vmod enter into a probe merit factor.
The electrostatic force terms of Equation 24 at 5fc and at 8fc are even further apart from the distance controller cutoff frequency. The term at 2fc does introduce some response of the distance controller, but this has a negligible effect on the Kelvin controller that demodulates at 4fc. Vice versa, the crosstalk of a topography varying at fc onto the VK voltage is a variation at 3fc damped by 24 dB:
Kelvin voltage noise PSD
The thermal noise PSD of the PLL frequency noise Equation 20 is constant and hence invariant under the frequency translation, yielding as Kelvin noise PSD, by dividing through the Kelvin gain Equation 26,
indicated as red curve in Figure 14. The integrated noise is
while the sensor noise of Equation 20 contributes to the Kelvin noise PSD
(blue curve of Figure 14). The integrated noise due to sensor noise, still with the condition that fmod = 4 BW, is
The following treatment supposes that one of the noise sources is dominant and hence the total integrated noise ΔVK is either equal to ΔVK,th or to ΔVK,S.
Merit factor and design optimization
We define the merit factor as
To obtain a merit factor, it is necessary to divide the reciprocal of the integrated noise ΔVK by the root of the effective probed surface Seff. It is obvious and a basic rule of statistics that a potential measurement on a n times bigger surface made in the same time with the same state of the art of measurement apparatus has a fluctuation of 1/ times the one on a simple surface.
We also divide by the AC voltage Vmod since for otherwise identical conditions, the Kelvin forward gain AK given by Equation 26 is proportional to it but at the same time this voltage has the effect of introducing an error on semiconductors by asymmetric band bending. The subject has been addressed by several authors [17-20]. If KFM is performed on a semiconductor, the AC bias applied to the tip causes a response of the underlying semiconductor that alternates between majority-carrier depletion and accumulation. The tip–substrate junction can be thought of as a capacitive voltage divider formed by the tip–substrate capacitance and the Mott–Schottky capacitance. We expect this description to be valid over a wide frequency. The competing process of inversion-layer buildup has a time constant that is typically on the order of seconds to minutes for industrial grade semiconductor and hence negligible even in FM-KFM. If charge capture and emission by defect states is involved, it is imaginable that time constants are such that frequency dependence or non-linearity can play a role. Due to the lack of detailed knowledge, we justify dividing the merit factor by Vmod.
The integrated noise ΔVK is dominated by thermal or detector noise depending on bandwidth and temperature. We define as crossover temperature Tcross the temperature above which at a given bandwidth, the integrated thermal noise of Equation 38 exceeds the integrated sensor noise Equation 40, while the design rule is respected:
Regarding the effective probed surface Seff, its absolute value is not known, but the relation between tip–sample distance and probed surface, as illustrated in Figure 16, is described by a power law derived from the second derivative of the capacitance [21]. Here we make the approximation that the probe oscillates with a small amplitude D0 around a larger average probe distance z.
Hence
For thermal noise domination, Equation 38, the merit factor is
If it is assumed that the maximum oscillation amplitude D0 cannot exceed a certain fraction of z and hence is proportional to it, it reduces to
Furthermore, a relation has to be respected between minimum tip–sample distance z and spring constant k to avoid snap to contact.
Figure 17 shows the tip in the attractive part of the van-der-Waals interaction. The force gradient in this field must not exceed the spring constant to avoid snap to contact. We take the attractive range of a Lennard-Jones type of potential
The force gradient is proportional to the second derivative:
To avoid snap to contact, the force gradient must be smaller than the cantilever stiffness
And hence Equation 46 reduces to
For comparison, a widely used merit factor for MEMS resonators is
and the one of minimum force detection is
This result, i.e., the maximization of f0Q/k0.69 is positioned between the usual MEMS benchmark f0Q and a merit factor f0Q/k found by Albrecht [3] for the minimum detectable force by noncontact AFM.
If the noise PSD is dominated by detector noise, Equation 40, then we obtain a merit factor MS instead of Equation 45:
similarly as above, D0 is a fraction of z and hence
Using Equation 49 for the relation between z and k yields
Unsurprisingly, for the case of dominating sensor noise, maximization of the merit factor requires minimizing the sensor noise. Both merit factors, Equation 50 and Equation 55, suggest downsizing both the probe spring constant and mass. If one considers f0 = , the exponents of k higher than 1/2 in the denominator yield increasing merit factors for decreasing stiffness. Both merit factors cannot be increased infinitely because downsizing the probe beyond a certain limit will decrease the Q-factor and increase sensor noise.
A table of merit factors for thermally dominated noise, sensor dominated noise and crossover criteria is given in Table 1. The table lists probe parameters, followed by a crossover criterion, Equation 42, the crossover temperature for a bandwidth of 50 Hz, the merit factor for dominant thermal noise according to Equation 50, and the merit factor for dominant detector noise according to Equation 55. For the stiffness of the Kolibri sensor, we use 1 MN/m, about the double of what is given in the documents from Specs [22]. The 540 kN/m is the spring constant of the entire needle which is suspended in the middle. In SPM operation, the two prongs are moving oppositely and the suspension remains stationary. Therefore, for comparison with the other probes, the effective stiffness of twice the given value has to be used. The use of the table for comparison of probe performance consists in first determining the crossover temperature as function of the desired bandwidth by multiplying the value Tcross/BW2 with BW2. If the working temperature is below the obtained crossover temperature, the merit factor MS applies and is obtained by dividing the value MS·BW3/2 through BW3/2. If the working temperature is above the crossover temperature, the merit factor Mth applies and is obtained by dividing Mth· by . The dominating merit factor among MS and Mth is the one with the lowest value, due to its definition containing the reciprocal of VK,n, according to Equation 41. The performance of probes with thermally dominated noise can be compared directly to others with dominating sensor noise.
Table 1: Key values, crossover criteria, and merit factors for different probes.
Cantilever (Nanosensors) | qPlus Sensor (Giessibl, 2011 [9]) | Kolibri Length Extension Resonator (Specs [22]) | IEMN Disk Resonator (Algre 2012 [23]) | |
---|---|---|---|---|
f0 (Hz) | 61836 | 30k | 1M | 1.45M |
Q | 22k8 | 200k | 14k | 1k6 |
k (N/m) | 1.2 | 1k8 | 1M | 135k |
Zn,S (m/) | 2e-13 | 6.2e-14 | 1e-15 | 1e-12 |
Tcross/BW2 (K/Hz2) | 6.81e-2 | 177.5 | 5.38e-2 | 594 |
Crossover Temp (K) at BW = 50 Hz | 170 | 444k | 135 | 1.48M |
Mth· | 1.71e16 | 2.83e15 | 4.89e14 | 3.97e14 |
Ms·BW3/2 | 6.75e16 | 2.05e14 | 1.97e15 | 1.56e13 |
The table shows that cryogenic cooling is useful only for reducing the thermal excitation of the Kolibri sensor and to some extent of cantilevers, whereas the qPlus and disk resonator have dominant detector noise at all achievable temperatures, recognizable by crossover temperatures in the kilo- or Mega-Kelvin range. (Detector noise was assumed temperature independent). The best FM-KFM performance is expected from standard cantilevers. It can be expected that these probes in combination with interferometric detection might benefit from cooling to temperatures even below liquid helium. Despite significant performance differences, the existence of all compared probe types seems to be justified. For instance, some environments require a need for electrical rather than optical deflection detection, and the performance criteria for topography imaging differ largely from the FM-KFM merit factor, due to the highly non-linear probe sample interaction that motivates a wide range of cantilever stiffness and oscillation amplitudes.
Absolute frequency limits irrespective of noise
The example treated here seems to have rather low performance compared to, e.g., video-rate SPM setups that claim to image biological processes in real-time (however in topography mode only). We emphasize that the choice of our bandwidth is our personal preference of making the compromise between bandwidth and noise. As stated above, the 30 Hz bandwidth leads to 22 mV signal fluctuation. Since the sampling circuitry has an anti-aliasing filter that cuts above half the sampling rate and it is not justified to smooth the image by slow response of the Kelvin and topography loop responses, we can acquire at 50 pixel per second, meaning that a line with 256 pixels is scanned back and forth in 10 seconds and an image at 256 × 256 resolution takes 45 minutes. We are used to acquire images with higher resolution over night. Since the output noise has been tracked down to thermal excitation and displacement detector noise, said compromise has universal validity. We also mention here that the exchange of the light source in the optical beam deflection sensor has already decreased the detector noise by an order of magnitude with respect to the original value, and that otherwise for the same choice of bandwidth, the detector noise would be dominating and the Kelvin signal fluctuation would be a multiple. In this short paragraph we address the question to what extent speed can be increased at the expense of noise and when other limitations apply.
- PLL bandwidth: for phase locked loops, the terms capture range and lock range denominate the frequency range in which the PLL can lock on to an incoming signal and maintain the phase lock. It is given as percentage of center frequency, depends on the degree of sophistication of the circuit (phase detection, filters) and is above 10 percent even for primitive monolithic circuits that use edge detection and simple filters such as the NE567 PLL tone decoder. The capture range is always below the lock range. The given percentage is the frequency shift of the frequency modulated signal, which is a function of both the excursion frequency and the modulation frequency. Without entering PLL theory in detail, we can say that a PLL bandwidth of 10 percent of the center frequency is realistic and it has been experimentally confirmed that our PLL bandwidth can be set to 5 kHz.
- AC frequency fmod and PLL bandwidth BWPLL: these frequencies have to be chosen such that fmod + BW ≤ BWPLL. Together with the design rule, for our example, we would obtain fmod = 4 kHz and BW = 1 kHz.
- Distance control: this component is probably the most limiting. In our setup where the sample is mounted on a 3 axis piezo scanner, the bandwidth is limited to a value between 100 and 200 Hz.
Consequently, if we had set the PLL bandwidth BWPLL = 5 kHz, the AC frequency fmod = 4 kHz and the bandwidth of the Kelvin controller BW = 1 kHz, according to Figure 9 the detector noise would be dominating and due to the power law with exponent 3/2 of Equation 40, we would expect Kelvin voltage fluctuation in the volt range. Furthermore, the distance control would not be able to keep up with the Kelvin loop.
Perspective on the ultimate probe and detector
It is obvious from the two merit factors that a reduction of both thermal and detector noise at the same time is difficult. If thermal excitation is dominant and the effort aims at reducing it, the frequency range where it dominates becomes smaller, as can be seen in Figure 9. Mass and spring constant cannot be reduced infinitely without reducing the Q-factor. Furthermore, increasing the merit factor in the thermally dominated case is a simple downsizing of the detector, and with the same type of sensor, would increase the sensor noise or decrease the Q-factor of the oscillator by sensor back-action (e.g., radiation pressure). Similarly, all attempts of improving the detector have a trend to increase invasiveness and to reduce the Q-factor. As long as one type of noise is clearly dominant, the remedy is to maximize the respective merit factor, keeping in mind the above dependencies. Present state of the art for measuring the excursion of harmonic oscillators consists in optical interferometry [24] or single electron transistors [25] used as position probe coupled to oscillators, combined with cooling of the resonator to cryogenic temperatures, possibly using laser cooling. These works aim at the Heisenberg limit and are not specific to scanning probe microscopy. Practical SPM systems seem to be still further away from the ultimate limit.
Conclusion
The dynamic behavior of an FM-KFM has been measured and modeled for a system with characteristics typically obtained in ultrahigh vacuum implementations. It has been shown that in a PLL based setup, the two main noise sources, thermal excitation and detector noise, transform into frequency noise exactly the same way as in a free-running oscillator, and that the PLL components do not contribute considerable noise, meaning that the main noise sources are sufficient to derive Kelvin voltage noise. Feedback parameters for PLL and Kelvin loop have been set for a stable behavior and been used for the numerical modeling of the noise propagation, yielding output noise spectra in agreement with the measurements. The choice of the AC modulation frequency to be four times the intended bandwidth has been proposed and justified as design rule. Based on the acquired knowledge, the KFM performance has been modeled for three other well-known AFM probes. A crossover criteria allows one to determine for each probe, depending on temperature, detector noise PSD, bandwidth and probe parameters, whether Kelvin output noise is dominated by thermal probe excitation or by detector noise. Depending on the regime, one of two merit factors apply to obtain the overall noise performance from instrument parameters, to suggest improvements and to allow for a comparison of different probes. Limitations to the optimization remain due to unresolved interdependent parameters, the trend of entering a thermally limited regime when improvement is made to detector noise and vice versa, and deteriorating one noise source when improving the other, ultimately merging into the uncertainty relation governing that a system cannot be measured without changing it by whatever kind of sensor back-action.
Experimental
The KFM is based on an Omicron ultrahigh vacuum variable temperature atomic force microscope (UHV-VT-AFM). It is operated by a Nanonis scanning probe microscopy (SPM) controller entirely based on digital signal processing (DSP). The probe that was used in these experiments is a platinum-iridium coated Nanosensors Point Probe Plus EFM tip with a spring constant between 1 and 3 N/m. Its resonance frequency f0 = 61.835 kHz and the Q-factor Q = 22800 were determined in vacuum by recording a resonance curve with the built in function of the Nanonis controller. The optical beam deflection detection uses a 20 mW Superluminescent (TM) light emitting diode that was operated at an intensity of 7 mW. About 0.5 mW intensity is received by the photodiode, which was estimated from its known current–intensity characteristics. To compensate the increased intensity of the light source, the gain of the transimpedance amplifier was reduced accordingly to avoid output voltage saturation. The sample is a gold coated silicon substrate (Omicron test sample). KFM measurements are performed while distance control is enabled.
Acknowledgements
This work was achieved with the financial support of the European Union, the French Government and the Regional Council, using the facilities of the EXCELSIOR Nanoscience Characterization Center. Further financial aid from the French Agence Nationale de Recherche (ANR-11-BS10-004 NANOKAN) and from a Merlion program of the French Embassy in Singapore (grant 8.11.08) is gratefully acknowledged. We thank Benjamin Walter for fruitful discussions and David Troadec for technical help.
References
-
Kikukawa, A.; Hosaka, S.; Imura, R. Appl. Phys. Lett. 1995, 66, 3510. doi:10.1063/1.113780
Return to citation in text: [1] -
Kitamura, S.; Iwatsuki, M. Appl. Phys. Lett. 1998, 72, 3154. doi:10.1063/1.121577
Return to citation in text: [1] -
Albrecht, T. R.; Grütter, P.; Horne, D.; Rugar, D. J. Appl. Phys. 1991, 69, 668. doi:10.1063/1.347347
Return to citation in text: [1] [2] [3] -
Glatzel, T.; Sadewasser, S.; Lux-Steiner, M. C. Appl. Surf. Sci. 2003, 210, 84. doi:10.1016/S0169-4332(02)01484-8
Return to citation in text: [1] -
Fukuma, T.; Kimura, M.; Kobayashi, K.; Matsushige, K.; Yamada, H. Rev. Sci. Instrum. 2005, 76, 53704. doi:10.1063/1.1896938
Return to citation in text: [1] [2] -
Kobayashi, K.; Yamada, H.; Matsushige, K. Rev. Sci. Instrum. 2009, 80, 043708. doi:10.1063/1.3120913
Return to citation in text: [1] [2] -
Polesel-Maris, J.; Venegas de la Cerda, M. A.; Martrou, D.; Gauthier, S. Phys. Rev. B 2009, 79, 235401. doi:10.1103/PhysRevB.79.235401
Return to citation in text: [1] [2] -
Diesinger, H.; Deresmes, D.; Nys, J.-P.; Mélin, T. Ultramicroscopy 2010, 110, 162. doi:10.1016/j.ultramic.2009.10.016
Return to citation in text: [1] -
Giessibl, F.-J.; Pielmeier, F.; Eguchi, T.; An, T.; Hasegawa, Y. Phys. Rev. B 2011, 84, 125409. doi:10.1103/PhysRevB.84.125409
Return to citation in text: [1] [2] [3] -
Lübbe, J.; Temmen, M.; Rode, S.; Rahe, P.; Kühnle, A.; Reichling, M. Beilstein J. Nanotechnol. 2013, 4, 32. doi:10.3762/bjnano.4.4
Return to citation in text: [1] -
Portes, L.; Ramonda, M.; Arinero, R.; Girard, P. Ultramicroscopy 2007, 107, 1027. doi:10.1016/j.ultramic.2007.03.012
Return to citation in text: [1] -
Giessibl, F.-J. Phys. Rev. B 1997, 56, 16010. doi:10.1103/PhysRevB.56.16010
Return to citation in text: [1] -
Rast, S.; Gysin, U.; Meyer, E. Phys. Rev. B 2009, 79, 054106. doi:10.1103/PhysRevB.79.054106
Return to citation in text: [1] -
Giessibl, F.-J. Rev. Mod. Phys. 2003, 75, 949. doi:10.1103/RevModPhys.75.949
Return to citation in text: [1] -
Colchero, J.; Cuenca, M.; Martínez, J.-F. G.; Abad, J.; García, B. P.; Palacios-Lidón, E.; Abellán, J. J. Appl. Phys. 2011, 109, 024310. doi:10.1063/1.3533769
Return to citation in text: [1] -
Gauthier, S. J. Appl. Phys. 2011, 110, 036107. doi:10.1063/1.3619776
Return to citation in text: [1] -
Leng, Y.; Williams, C. C.; Su, L. C.; Stringfellow, G. B. Appl. Phys. Lett. 1995, 66, 1264. doi:10.1063/1.113257
Return to citation in text: [1] -
Henning, A.-K.; Hochwitz, T. Mater. Sci. Eng. B 1996, 42, 88. doi:10.1016/S0921-5107(96)01688-1
Return to citation in text: [1] -
Sommerhalter, C.; Matthes, T. W.; Glatzel, T.; Jäger-Waldau, A.; Lux-Steiner, M. C. Appl. Phys. Lett. 1999, 75, 286. doi:10.1063/1.124357
Return to citation in text: [1] -
Sommerhalter, C.; Glatzel, T.; Matthes, T.; Jager-Waldau, A.; Lux-Steiner, M. Appl. Surf. Sci. 2000, 157, 263. doi:10.1016/S0169-4332(99)00537-1
Return to citation in text: [1] -
Mélin, T.; Diesinger, H.; Deresmes, D.; Stiévenard, D. Phys. Rev. B 2004, 69, 35321. doi:10.1103/PhysRevB.69.035321
Return to citation in text: [1] -
Specs, Kolibri Sensor, Advanced Quartz Sensor Technology for NC-AFM. http://www.specs.de/cms/front_content.php?idcat=246.
Return to citation in text: [1] [2] -
Algre, E.; Xiong, Z.; Faucher, M.; Walter, B.; Buchaillot, L.; Legrand, B. J. Microelectromech. Syst. 2012, 21, 385. doi:10.1109/JMEMS.2011.2179012
Return to citation in text: [1] -
Schliesser, A.; Arcizet, O.; Rivière, R.; Anetsberger, G.; Kippenberg, T.-J. Nat. Phys. 2009, 5, 509. doi:10.1038/nphys1304
Return to citation in text: [1] -
LaHaye, M.-D.; Buu, O.; Camarota, B.; Schwab, K.-C. Science 2004, 304, 74. doi:10.1126/science.1094419
Return to citation in text: [1]
25. | LaHaye, M.-D.; Buu, O.; Camarota, B.; Schwab, K.-C. Science 2004, 304, 74. doi:10.1126/science.1094419 |
1. | Kikukawa, A.; Hosaka, S.; Imura, R. Appl. Phys. Lett. 1995, 66, 3510. doi:10.1063/1.113780 |
3. | Albrecht, T. R.; Grütter, P.; Horne, D.; Rugar, D. J. Appl. Phys. 1991, 69, 668. doi:10.1063/1.347347 |
5. | Fukuma, T.; Kimura, M.; Kobayashi, K.; Matsushige, K.; Yamada, H. Rev. Sci. Instrum. 2005, 76, 53704. doi:10.1063/1.1896938 |
14. | Giessibl, F.-J. Rev. Mod. Phys. 2003, 75, 949. doi:10.1103/RevModPhys.75.949 |
4. | Glatzel, T.; Sadewasser, S.; Lux-Steiner, M. C. Appl. Surf. Sci. 2003, 210, 84. doi:10.1016/S0169-4332(02)01484-8 |
15. | Colchero, J.; Cuenca, M.; Martínez, J.-F. G.; Abad, J.; García, B. P.; Palacios-Lidón, E.; Abellán, J. J. Appl. Phys. 2011, 109, 024310. doi:10.1063/1.3533769 |
16. | Gauthier, S. J. Appl. Phys. 2011, 110, 036107. doi:10.1063/1.3619776 |
3. | Albrecht, T. R.; Grütter, P.; Horne, D.; Rugar, D. J. Appl. Phys. 1991, 69, 668. doi:10.1063/1.347347 |
2. | Kitamura, S.; Iwatsuki, M. Appl. Phys. Lett. 1998, 72, 3154. doi:10.1063/1.121577 |
13. | Rast, S.; Gysin, U.; Meyer, E. Phys. Rev. B 2009, 79, 054106. doi:10.1103/PhysRevB.79.054106 |
8. | Diesinger, H.; Deresmes, D.; Nys, J.-P.; Mélin, T. Ultramicroscopy 2010, 110, 162. doi:10.1016/j.ultramic.2009.10.016 |
10. | Lübbe, J.; Temmen, M.; Rode, S.; Rahe, P.; Kühnle, A.; Reichling, M. Beilstein J. Nanotechnol. 2013, 4, 32. doi:10.3762/bjnano.4.4 |
7. | Polesel-Maris, J.; Venegas de la Cerda, M. A.; Martrou, D.; Gauthier, S. Phys. Rev. B 2009, 79, 235401. doi:10.1103/PhysRevB.79.235401 |
11. | Portes, L.; Ramonda, M.; Arinero, R.; Girard, P. Ultramicroscopy 2007, 107, 1027. doi:10.1016/j.ultramic.2007.03.012 |
6. | Kobayashi, K.; Yamada, H.; Matsushige, K. Rev. Sci. Instrum. 2009, 80, 043708. doi:10.1063/1.3120913 |
5. | Fukuma, T.; Kimura, M.; Kobayashi, K.; Matsushige, K.; Yamada, H. Rev. Sci. Instrum. 2005, 76, 53704. doi:10.1063/1.1896938 |
9. | Giessibl, F.-J.; Pielmeier, F.; Eguchi, T.; An, T.; Hasegawa, Y. Phys. Rev. B 2011, 84, 125409. doi:10.1103/PhysRevB.84.125409 |
7. | Polesel-Maris, J.; Venegas de la Cerda, M. A.; Martrou, D.; Gauthier, S. Phys. Rev. B 2009, 79, 235401. doi:10.1103/PhysRevB.79.235401 |
6. | Kobayashi, K.; Yamada, H.; Matsushige, K. Rev. Sci. Instrum. 2009, 80, 043708. doi:10.1063/1.3120913 |
9. | Giessibl, F.-J.; Pielmeier, F.; Eguchi, T.; An, T.; Hasegawa, Y. Phys. Rev. B 2011, 84, 125409. doi:10.1103/PhysRevB.84.125409 |
23. | Algre, E.; Xiong, Z.; Faucher, M.; Walter, B.; Buchaillot, L.; Legrand, B. J. Microelectromech. Syst. 2012, 21, 385. doi:10.1109/JMEMS.2011.2179012 |
24. | Schliesser, A.; Arcizet, O.; Rivière, R.; Anetsberger, G.; Kippenberg, T.-J. Nat. Phys. 2009, 5, 509. doi:10.1038/nphys1304 |
9. | Giessibl, F.-J.; Pielmeier, F.; Eguchi, T.; An, T.; Hasegawa, Y. Phys. Rev. B 2011, 84, 125409. doi:10.1103/PhysRevB.84.125409 |
22. | Specs, Kolibri Sensor, Advanced Quartz Sensor Technology for NC-AFM. http://www.specs.de/cms/front_content.php?idcat=246. |
3. | Albrecht, T. R.; Grütter, P.; Horne, D.; Rugar, D. J. Appl. Phys. 1991, 69, 668. doi:10.1063/1.347347 |
22. | Specs, Kolibri Sensor, Advanced Quartz Sensor Technology for NC-AFM. http://www.specs.de/cms/front_content.php?idcat=246. |
17. | Leng, Y.; Williams, C. C.; Su, L. C.; Stringfellow, G. B. Appl. Phys. Lett. 1995, 66, 1264. doi:10.1063/1.113257 |
18. | Henning, A.-K.; Hochwitz, T. Mater. Sci. Eng. B 1996, 42, 88. doi:10.1016/S0921-5107(96)01688-1 |
19. | Sommerhalter, C.; Matthes, T. W.; Glatzel, T.; Jäger-Waldau, A.; Lux-Steiner, M. C. Appl. Phys. Lett. 1999, 75, 286. doi:10.1063/1.124357 |
20. | Sommerhalter, C.; Glatzel, T.; Matthes, T.; Jager-Waldau, A.; Lux-Steiner, M. Appl. Surf. Sci. 2000, 157, 263. doi:10.1016/S0169-4332(99)00537-1 |
21. | Mélin, T.; Diesinger, H.; Deresmes, D.; Stiévenard, D. Phys. Rev. B 2004, 69, 35321. doi:10.1103/PhysRevB.69.035321 |
© 2014 Diesinger et al; licensee Beilstein-Institut.
This is an Open Access article under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
The license is subject to the Beilstein Journal of Nanotechnology terms and conditions: (http://www.beilstein-journals.org/bjnano)