Unlocking higher harmonics in atomic force microscopy with gentle interactions
‡,1,
‡,1,
1 and
2,3
Sergio Santos
Departament de Disseny i Programació de Sistemes Electrònics, UPC - Universitat Politècnica de Catalunya Av. Bases, 61, 08242 Manresa (Barcelona), Spain
Departament de Disseny i Programació de Sistemes Electrònics, UPC - Universitat Politècnica de Catalunya Av. Bases, 61, 08242 Manresa (Barcelona), Spain
Departament de Disseny i Programació de Sistemes Electrònics, UPC - Universitat Politècnica de Catalunya Av. Bases, 61, 08242 Manresa (Barcelona), Spain
1Departament de Disseny i Programació de Sistemes Electrònics, UPC - Universitat Politècnica de Catalunya Av. Bases, 61, 08242 Manresa (Barcelona), Spain
2ICN2 - Institut Catala de Nanociencia i Nanotecnologia, Campus UAB, 08193 Bellaterra (Barcelona), Spain
3CSIC - Consejo Superior de Investigaciones Cientificas, ICN2 Building ,08193 Bellaterra (Barcelona), Spain
Guest Editors: U. D. Schwarz and M. Z. Baykara Beilstein J. Nanotechnol.2014,5, 268–277.https://doi.org/10.3762/bjnano.5.29 Received 09 Oct 2013,
Accepted 14 Feb 2014,
Published 11 Mar 2014
In dynamic atomic force microscopy, nanoscale properties are encoded in the higher harmonics. Nevertheless, when gentle interactions and minimal invasiveness are required, these harmonics are typically undetectable. Here, we propose to externally drive an arbitrary number of exact higher harmonics above the noise level. In this way, multiple contrast channels that are sensitive to compositional variations are made accessible. Numerical integration of the equation of motion shows that the external introduction of exact harmonic frequencies does not compromise the fundamental frequency. Thermal fluctuations are also considered within the detection bandwidth of interest and discussed in terms of higher-harmonic phase contrast in the presence and absence of an external excitation of higher harmonics. Higher harmonic phase shifts further provide the means to directly decouple the true topography from that induced by compositional heterogeneity.
It has long been recognized in the community that higher harmonics encode detailed information about the non-linearities of the tip–sample interaction in dynamic atomic force microscopy (AFM) [1-5]. Physically, non-linearities relate to the chemical and mechanical composition [6] of the tip–sample system and imply that higher harmonics can be translated into conservative and dissipative [7] nanoscale and atomic properties [8]. Furthermore, conventional dynamic AFM can already reach molecular [9,10], sub-molecular [11] and atomic [12,13] resolution in some systems. Thus, the simultaneous detection and interpretation of multiple higher harmonic signals while scanning [14] can lead to spectroscopy-like capabilities [15,16], such as chemical identification, with similar or higher resolution [5,17,18]. The higher harmonic approach however, and particularly in other than highly damped environments [19,20], requires dealing with the recurrent challenge of detecting higher harmonics [1,3,21,22]. Higher harmonics are a result of the non-linear tip–sample interaction in the sense that the interaction effectively acts as the driving force of each harmonic component [7]. Accordingly, relatively high peak forces, of the order of 1–100 nN, are required [22,23] to excite higher harmonics above the noise level. In order to address this issue, in 2004 Rodriguez and García [23] proposed to drive the second higher flexural mode of the cantilever with an external drive. In this way, and by driving with sufficiently small (sub-nanometer) second mode amplitudes, the first mode amplitude [24] or frequency [17] can be employed to track the sample in amplitude or frequency modulation (AM and FM), respectively. The second mode can then be left as an open loop for high sensitivity mapping of compositional variations [25] or as a closed loop, in which case the tip–sample stiffness kts can be computed [17,26]. More recently, the multifrequency AFM approach has been extended to employ three flexural modes [27] and/or simultaneous torsional modes [28], for which, typically, the frequency and mode under consideration are externally excited [24]. In summary, FM and/or AM feedback systems can be employed in one [29], several [27] or all of the modes under consideration in order to quantify properties on the nanoscale through observables [30] while simultaneously enhancing sensitivity and throughput [31]. The dynamics in the multifrequency approach, however, might lead to extra complexities in the analysis, acquisition and interpretation of data [31,32]. For example, recent studies [31] show that multiple regimes of operation might follow depending on the relative kinetic energy between the higher mode of choice and the fundamental eigenmode [31,33].
Here, exact multiple harmonics of the fundamental drive frequency are externally excited above the noise level to open multiple contrast channels that are sensitive to compositional variations. The focus is on amplitude modulation (AM) AFM, in which the fundamental amplitude A1 ≡ A tracks the sample as usual. For standard cantilevers the eigenmodes are nonharmonic [29]. That is, the natural resonant frequencies of the cantilevers are not integer multiples. Furthermore, these natural frequencies relate to the geometry and mechanical properties of the cantilever [34]. The practical implication is that it is only easy to induce large oscillations at the frequencies that coincide with these natural frequencies. Nevertheless the tip–sample coupling always occurs via harmonic frequencies. This is because a periodic motion always implies that there is a fundamental frequency and that all other higher frequencies are integer multiples of the fundamental [35]. The implication is that externally introducing frequencies other than harmonic frequencies could induce a fundamental sub-harmonic frequency [24,35]. In short, the incommensurability between external drives in the standard multifrequency approach implies that the cantilever motion is not exactly periodic relative to the fundamental drive and that a sub-harmonic excitation typically follows [32]. Furthermore, simplifications in eigenmode frequency shift theory [36] might lead to inconsistencies [37]. This issue becomes more prominent when dealing with third or higher eigenmodes [27,38], for which the theory is now emerging [31]. The introduction of exact harmonic external drives keeps the fundamental frequency intact and the analytical expressions are simplified by orthogonality. Furthermore 2(N−1) observables, i.e., higher harmonic amplitudes and phases, are made available even with peak forces no higher than 200 pN, as they are required [25,39] for high resolution and minimally invasive imaging of soft matter. Thermal fluctuations are also considered here in order to establish a possible loss of contrast due to fundamental sources of noise. It is also shown that true topography and apparent topography, which is induced by chemical heterogeneity, can be decoupled at once by monitoring the phase contrast of higher harmonics.
Results and Discussion
Consider the equation of motion of the mth eigenmode
where k(m), Q(m), ω(m), and z(m) are the spring constant, quality factor, natural frequency and position of the mth eigenmode. The term FD stands for the external driving force
where the subscript without brackets, n, indicates the harmonic number. Note that here ωn = nω, where ω is the fundamental drive frequency set near mode m = 1, i.e., ω = ω1 ≡ ω(1). The term Fts is the tip–sample force, which is a function of both the tip–sample distance, d, and velocity, . Here however, we focus on conservative forces since these are present even with gentle interactions. Hence we can write Fts(d). Since the higher harmonic amplitudes here are externally excited, the number of harmonics N that is to be monitored can, in principle, be arbitrarily chosen up to the limits of frequency detection, i.e., of the order of MHz, without compromising detection. The main constraint is that the number of higher modes, M, that is to be considered needs to be consistent with the number of higher harmonics N that are to be analysed [22]. For simplicity, we consider M = 2 and N = 10 in the numerical analysis without loss of generality. For clarity we emphasize that M is the number of modes and N is the number of harmonics taken into consideration in the analysis in this work. A particular mode or harmonic is referred to in lower case, i.e., m or n respectively.
where 1 is assumed when no subscripts are given. The relationships (ω/ω(m))2 = k/k(m) and Q/Q(m) = ω/ω(m)[7] have been employed in Equation 4 and it has been assumed that the fundamental drive frequency ω is set near ω(1). Furthermore, in Equation 4A(m)n and An are the amplitudes of the nth harmonic that correspond to the position of mode m, i.e., z(m), and to the absolute position of the tip, i.e., z, respectively. Also
where and are the phase shifts of the nth harmonic that correspond to the mth mode position and the absolute position, z, respectively, and En is the energy involved with the nth harmonic tip–sample interaction. Near the modal frequency ω(m) only the mth mode significantly contributes to the interaction and B(m)n ≈ 0 and C(m)n ≈ 1 in Equation 4. This approximation has been currently employed in the literature [6]. Nevertheless, far from the modes, these terms might not be zero. To allow for simple analytical formulae and ease the qualitative interpretation we consider the harmonics close to the modes only [6]. Then
If the nth drive F0n is zero, then
Equation 9 is the energy transferred to the nth harmonic of the cantilever through the tip–sample interaction. It should be noted that this is consistent with a conservative tip–sample force Fts(d) since the energy is provided during each cycle by the external driving force(s). The quadratic dependence of the energy En on nAn is of particular relevance for the detection of higher harmonics. First, Equation 9 implies that for a given amplitude An the transfer of energy En scales quadratically with the harmonic number. This explains why for sufficiently large n, higher harmonics are typically undetectable. Second, the proportionality between En and in Equation 9 explains why for higher harmonic amplitudes to be detected, the interaction in Equation 7 needs to be considerably large, even when n is not necessarily very large.
From Equation 8 it follows that An can be set to any arbitrary value by increasing F0n, even if there is no tip–sample energy transfer, i.e., En = 0. The higher harmonics for the free cantilever are termed A0n. This case corresponds to a free cantilever oscillating sufficiently high above the sample (A/A0 = 1) as illustrated in Figure 1 (circles). The data has been acquired by numerically solving the simultaneous equations in Equation 1 for the first two flexural modes, i.e., M = 2, and for N = 10. Furthermore, since only long range attractive forces are of interest here, the tip–sample force is simply [23]
where R is the tip radius, H is the Hamaker constant and a0 is an intermolecular distance (a0 = 0.165 nm throughout and in all the data here, we consider d > a0 throughout). It is relevant to note that the Hamaker constant depends on the tip and sample in the sense that its value is determined by the atomic composition or chemical elements that compose the tip and the sample [40,41]. For this reason, in this work we will employ the terms chemistry, Hamaker and tip–sample composition or chemistry interchangeably. The common parameters in this work are k = 2 N/m, Q = 100, ω = 2π·70 kHz and R = 7 nm, i.e., they correspond to commercially available standard probes for AM AFM. Furthermore, in Figure 1, H = 6.2 × 10−19 J, i.e., it is close to that calculated for materials such as polystyrene or fused quartz [40]. The parameters for the second mode have been obtained with the above formulae [7]. The modal frequencies 1 and 2 are shown with dashed lines. The phase shifts are shown in the vertical axis in Figure 1 for each harmonic. The actual harmonic amplitudes An that resulted when interacting are not shown, instead An ≈ A0n is given throughout. The case of a free cantilever (circles) shows that the fundamental phase shift is exactly 90 degrees as expected while the higher harmonic phase shifts (n > 1) lie either close to 180° or to 0°. This is in agreement with Equation 8 when En ≈ 0 since then
where the approximation F0n ≈ k(m)n2A0n (near m) has been employed. Also from Figure 1 (circles) it follows that for a free cantilever, and when n is higher than the modal frequency (close to a given mode and for n > 1), ≈ 180°. When n is lower than the modal frequency ≈ 0°. This is true irrespective of the value of A0n. When the tip is allowed to interact with the sample En ≠ 0 and, from Equation 8, the phase shift is affected by the interaction. Nevertheless, the weight of the driving force, i.e., the first term in Equation 8, increases with increasing F0n, or A0n, and then the sensitivity of to En might be compromised. This is confirmed in Figure 1 by allowing a gentle interaction, i.e., A01 ≡ A0 = 4 nm and A/A0 ≈ 0.9 (also Figure 2 and Figure 3), and monitoring when A0n = 1 pm (squares) and A0n = 100 pm (triangles). When A0n = 100 pm (triangles) all remain close to 180° or 0°. A shift in phase, i.e., from 180° to 0°, is observed for n = 2 only. While these jumps of nearly 180° might be of interest they are ignored from now on. The reader can refer to recent works that discuss multiple regimes of operation in bimodal AFM [31,33]. It follows that variations in Hamaker are not detected by higher harmonic frequencies when A0n = 100 pm. When A0n = 1 pm (squares), however, the values of are not exactly 180° or 0° for some n. Thus, the values are now sensitive to the Hamaker values or tip–sample forces. The peak forces were 140 pN (circles) and 160 pN (triangles) respectively.
The loss of phase sensitivity to Hamaker variations with increasing A0n is further corroborated with the use of Figure 2 and by varying the Hamaker values from H1 = 0.2 × 10−19 J to H2 = 1.4 × 10−19 J, and setting A0n = 1 pm (circles), A0n =10 pm (squares) and A0n = 100 pm (triangles). This range of H is characteristic of materials interacting in ambient conditions [40]. The y-axis stands for the contrast in higher harmonic phase Δ = abs((H2) − (H1)). We consider that variations, for which Δ > 0.2° lie above the noise of the instrument and can potentially be detected. The corresponding variations in peak forces were 63, 47 and 79 pN respectively. The sensitivity of Δ is clearly controlled by the chosen values of A0n. For example, if A0n = 100 pm then Δ < 0.2° throughout. If A0n = 1 or 10 pm, however, then Δ > 0.2° at least for some n. In particular, if A0n = 1 pm then Δ > 0.2° for all n. This implies that all the externally excited higher harmonics act as simultaneous contrast channels that are sensitive to Hamaker, or chemical, variations.
In Figure 3 the sensitivity of Δ when A0n = 1 pm is tested by varying H (a) from H1 = 0.2 × 10−19 J to H2 = 0.4 × 10−19 J (peak force variation of 29 pN, circles), (b) from H1 = 0.6 × 10−19 J to H2 = 0.8 × 10−19 J (peak force variation of 8 pN, squares) and (c) from H1 = 1.2 × 10−19 J to H2 = 1.4 × 10−19 J (peak force variation of 3 pN, triangles). The shifts Δ are larger than 0.2° for all n provided the variations in peak force are large enough (circles). If the variations in the peak force are sufficiently small then Δ > 0.2° for some n only. Also, it can be deduced by inspection that, in general, Δ escalates with variations in peak force and changes non-linearly with variations in Hamaker since H2 − H1 = 0.2 × 10−19 J throughout in the figure. In fact, from Figure 3, the total contributions to the phase shift calculated as the sums ΣΔ (n = 1–9) are 119.8, 19.3 and 5.4° and decrease with decreasing the variations in peak force, i.e., 29, 8 and 3 pN, respectively.
It is also interesting to note that the source of variations in peak force with variations in Hamaker H (Equation 10), i.e., van der Waals forces, relates to variations in the distance of minimum approach, dm, with variations in H. To be more specific, dm, increases with increasing H. For example, in the simulations, by varying H from H1 = 0.2 × 10−19 J to H2 = 1.4 × 10−19 J the variation is Δdm ≈ 0.83 nm. This would experimentally result in a chemistry-induced apparent topography of approximately Δzc ≈ 0.83 nm. In standard AM AFM, in which a single frequency is externally excited, this apparent topography cannot be distinguished from true topography in the presence of conservative forces only (Figure 4). A true topography can only be reconstructed from AM AFM results, if there is a variation in topography only (Figure 4a). This means that the composition of the sample is homogeneous throughout. In particular, the above discussion indicates that variations in H, or chemistry alone, produce variations in apparent topography in AM AFM, for which Δzc > 0 nm (Figure 4b). The excitation of higher harmonics, however, provides experimental observables to differentiate between the two cases. Namely, the true reconstructed topography results only if Δ = 0° for all n. That is, if Δ > 0°, even for a single n, there is a contribution to apparent topography induced by chemistry or other compositional variations.
Thermal noise and higher harmonic external drives
As stated in the introduction, it has long been known that under ambient conditions higher harmonic amplitudes might be too small to be detected [3,15,42]. This is particularly true when monitoring higher harmonics and simultaneously applying gentle tip–sample forces [23]. In liquid environments, however, the second harmonic amplitude might be large enough [43] to be recorded to map the properties even of living cells [44,45]. Still, even in highly damped environments, harmonic amplitudes rapidly decrease with increasing harmonic number particularly when imaging soft matter [6,15,46]. The main discussion above has focused on externally driving higher harmonics to amplitudes that could be experimentally detected. Then, once these amplitudes are sufficiently high, the phase shifts Δ have been employed to map the composition through variations in the tip–sample Hamaker constant, H, in Equation 10. In this section, the presence of thermal noise is discussed with respect to the contrast in amplitude ΔAn and phase Δ in the presence and absence of external drive forces at the higher harmonics frequencies.
First an example of the magnitude of the harmonic amplitudes and respective phase shifts that would result when higher harmonics are not externally excited is given (Table 1). In order to sense long-range forces only, the cantilever is driven with relatively small amplitudes, i.e., A0 = 4 nm and A/A0 = 0.9 as in the examples above. The harmonic amplitudes An are given in pm. Two examples for the amplitude response are shown, one for amplitudes resulting from H1 = 0.2 × 10−19 J (top row) and one for H2 = 1.4 × 10−19 J (second row). For H1, A2 is approx. 4 pm whereas A3 and A6 are approx. 1 pm. All other higher harmonics lie below 1 pm. For H2, A2 is approx. 3 pm and all other higher harmonics have values below 1 pm. The difference in amplitudes ΔAn = An(H2) − An(H1) that results from the variation in H is also given in the table. Only the second harmonic results in variations above 1 pm. Practically, these results imply that while higher harmonic amplitudes depend on the value of the Hamaker constant, or sample composition, the amplitude values are typically in the order of 1 pm or fractions of a pm. This is also true for variations in higher harmonic amplitudes ΔAn. The corresponding phase shifts and variations in phase shifts Δ are also shown in Table 1 for H1 and H2. These are of the order of a hundredth of a degree or less except for sufficiently high harmonic numbers, i.e., n = 9 and 10. The amplitudes for these higher harmonics, however, are of the order of tens of femtometers or less.
Table 1:
Harmonic amplitudes An and the corresponding phase shifts that result from Hamaker values of H1 = 0.2 × 10−19 J and H2 = 1.4 × 10−19 J. The differences in amplitudes ΔAn and phases Δ are also shown. A single external drive force has been employed (the fundamental frequency) and no thermal noise has been allowed.
A1
A2
A3
A4
A5
A6
A7
A8
A9
A10
An [pm] for H1
3600.00
4.38
1.03
0.21
0.16
1.11
0.39
0.12
0.06
0.03
An [pm] for H2
3600.00
3.31
0.57
0.08
0.05
0.23
0.06
0.01
0.00
0.00
ΔA1
ΔA2
ΔA3
ΔA4
ΔA5
ΔA6
ΔA7
ΔA8
ΔA9
ΔA10
ΔAn [pm]
0.00
−1.07
−0.46
−0.12
−0.12
−0.89
−0.33
−0.11
−0.05
−0.03
[°] for H1
115.83
141.25
167.20
12.77
39.91
66.18
90.49
116.55
142.44
168.29
[°] for H2
115.84
141.25
167.20
12.77
39.90
66.18
90.48
116.52
142.29
167.78
Δ
Δ
Δ
Δ
Δ
Δ
Δ
Δ
Δ
Δ
Δ [°]
0.00
0.00
0.00
0.00
−0.01
0.00
0.00
−0.03
−0.14
−0.50
Thermal fluctuations are a fundamental source of intrinsic noise in atomic force microscopy [47]. Thus, while other sources of intrinsic and extrinsic noise should be acknowledged and might be present in a given experiment, thermal fluctuations are analyzed next in terms of their effects on amplitude and phase shifts. This should provide a measure of the impact of thermal noise on the enhanced contrast reported in this work (Figure 2 and Figure 3). Other technical issues such as tilt and probe geometry have also been ignored for simplicity since these typically involve a correction factor [48]. As in the work of Butt and Jaschke [47], the equipartition theorem is employed to estimate the thermal noise present in a given mode. However, since higher harmonics are discussed here, particular emphasis should be given to the noise at the frequencies of interest, i.e., at exact harmonic frequencies, and the noise in the detection bandwidths of interest. Then, the thermal noise power ΔPTN(Δf) in the detection bandwidth of interest, Δf, can be defined as
where TN stands for thermal noise, fn is the frequency of interest (ωn = 2πfn), that is the frequency of a particular harmonic n, GTN is the power spectral density due to thermal noise, and |HZF|2 is the modulus of the squared transfer function of a particular mode m of position zm relative to thermal force FTN. If GTN is assumed to be constant for the bandwidth of interest in AFM experiments, i.e., f = 102–106, it follows from Equation 1 that the thermal energy in a given mode m, by invoking the equipartition theorem, is
where here T = 300 K throughout, f(m) is the natural resonant frequency of mode m in Hz and df = (f(m)/ω(m))dω. Then
From Equation 13 and Equation 14, the thermal noise power in the detection bandwidth of interest, ΔPTN(Δf), is found to be
Finally, the associated amplitude due to thermal noise ATN in the detection bandwidth Δf is
It should be noted that ATN gives the contribution of thermal noise to the amplitude of a given mode m only. Each modal contribution of thermal noise to the amplitude should be calculated separately for each frequency in the formalism developed here. A driving force, FTN, can also be associated to thermal noise and the respective amplitude, ATN, (Equation 16) through a standard expression [49]
Equation 17 gives the effective drive force FTN due to thermal fluctuations that should be expected for a given detection bandwidth Δf and a given mode m. Since the upper boundaries for noise will be considered here, the phase of the thermal noise signal has been set to be in quadrature with respect to the external drive, i.e., either the fundamental external drive or the higher harmonic external drives when these are present. Focus is now placed on the harmonics n = 1, 2, 3 (close to the fundamental frequency of mode 1) and 6 (close to the fundamental frequency of mode 2), since these are sufficiently close to a given mode that only the contribution of thermal noise to the amplitude from a single mode needs to be considered. This simplifies the following discussion.
In Table 2 the amplitudes ATNn and forces FTNn calculated for three different values of detection bandwidth Δf (5 kHz, 2 kHz and 0.2 kHz) are shown for n = 1, 2, 3 and 6. The values have been computed with the use of Equation 16 and Equation 17, with frequencies centered at the harmonic frequencies fn, for a given detection bandwidth Δf. It is interesting to note that ATN1 lies between 44 and 19 pm for the three choices of detection bandwidth. These values are in agreement with those expected from an analysis that implies that all the thermal noise is centered exactly at resonance [47]. This is because the Q factors are relatively high (Q1 = 100 and Q2 = 600). The values of the thermal-noise amplitude expected at harmonics 2, 3 and 6 however are of the order of 0.1–1.0 pm.
Table 2:
Amplitudes ATNn resulting from thermal noise for n = 1, 2, 3 and 6 and respective drive forces FTNn for detection bandwidths Δf of 5, 2 and 0.2 kHz.
Δf [kHz]
ATN1 [pm]
FTN1 [pN]
ATN2 [pm]
FTN2 [pN]
ATN3 [pm]
FTN3 [pN]
ATN6 [pm]
FTN6 [pN]
5
62.23
1.27
0.42
2.83
0.17
2.83
0.42
2.83
2
56.57
1.13
0.28
1.70
0.14
1.70
0.28
1.70
0.2
26.87
0.57
0.08
0.57
0.04
0.57
0.07
0.57
The effects that the thermal noise amplitudes in Table 2 have on the enhanced contrast reported in this work have been analyzed by adding the associated thermal noise forces, also shown in Table 2, to the equation of motion in Equation 1. The discussion below focuses on the values obtained for Δf = 2 kHz in Table 2 since this is a detection bandwidth of practical relevance in standard AFM experiments [50].
The sensitivity of the phase shift to noise and signal can be defined here, and for the purpose of phase shifts in AM AFM, as follows. First assume that noise is allowed according to Table 2 (Δf = 2 kHz) for a given value of the Hamaker constant, H. Here both H1 = 1.4 × 10−19 J and H2 = 1.4 × 10−19 J have been used in the simulations. According to this, thermal noise alone should lead to a difference in phase shift Δ(H) = (ATN > 0) − (ATN = 0) for a given value of H since there is an effective driving force FTNn due to thermal fluctuations (Table 2). The average of Δ for the two Hamaker values can be taken as the noise in the phase signal as follows
where TN stands for thermal noise as usual and Δ(TN) stands for the difference in phase shift at harmonic n that induced by thermal noise alone. Next the signal is defined as the phase shift induced by variations in Hamaker alone
Finally, a parameter that quantifies the sensitivity of the phase shift to noise and signal, the phase ratio PR() can be defined from the ratio between Equation 19 and Equation 18:
Large values of PR result in a high sensitivity of the phase shift to the signal, whereas low values of PR indicate a sensitivity of the phase shift to noise only. Three cases are discussed, which, for simplicity, focus on harmonics 2, 3 and 6 only and on A0n = 0, 1 and 10 pm.
Case 1: First, no higher harmonic external drives are allowed, which implies that A0n = 0 in Equation 2 for n > 1. This is the standard operational mode in dynamic AFM, in which a single external drive is employed. In this case we have PR = 0 throughout (Table 3).
Case 2: Higher harmonic external drives are allowed. In particular, A0n = 1 pm in Equation 2 for n > 1. This is the proposed mode of operation in this work. In this case we have PR > 1 throughout but the exact value depends on harmonic number (Table 3).
Case 3: Higher harmonic external drives are allowed. In particular, A0n = 10 pm in Equation 2 for n > 1. This is the proposed mode of operation in this work. When compared to case 2, however, the magnitudes of the external drives have been increased. In this case we also have PR > 1 throughout (Table 3).
Table 3:
The phase ratio for a given harmonic phase shift n, PR(), as defined by Equation 20 when 1) no higher harmonic external drives are allowed (A0n = 0) and when external drives lead to 2) A0n = 1 pm and 3) A0n = 10 pm.
PR ()
PR ()
PR ()
case 1: A0n =0
0.00
0.00
0.00
case 2: A0n =1 pm
1.90
22.09
7.29
case 3: A0n =10 pm
5.20
2.01
195.85
When looking at Table 3, one should recall that these are the upper-boundary values for noise since the phase of the thermal noise drives was set to be in quadrature. In summary, Table 3 shows that the phase ratio PR increases when external drives are applied at a given exact harmonic frequency, i.e., when A0n > 0. This is consistent with standard multifrequency operation, for which impressive results have already been obtained by exciting frequencies close to the resonant frequency of the second flexural mode [17,25,26]. In standard monomodal dynamic AFM, in which a single external drive is employed, the higher harmonics are excited by the tip–sample interaction according to Equation 9. That is, energy needs to flow into the higher harmonic frequencies in order to increase the amplitude signal. It is reasonable to assume that the increase in the sensitivity of the phase shift to the signal, i.e., the force, when external drives are applied is a consequence of energy both entering and leaving the given harmonic frequency of choice. That is, the fact that energy is supplied by the external drive at a given harmonic n implies that both positive and negative energy transfer might also occur at that frequency. Furthermore, when external drives are employed, this transfer occurs for a given phase shift that is now measured relative to the angle of the driving force. This is in agreement with the presence of the phase shift in Equation 8 and the absence of the phase shift in Equation 9 and might be related to the increase in the sensitivity of the phase shift to the tip–sample force as predicted here.
Conclusion
In summary, we have introduced a method that makes readily accessible an arbitrary number of exact higher harmonics by externally driving them with amplitudes above the noise level. Driving with exact higher harmonics does not introduce sub-harmonic frequencies to the motion and the amplitudes do not significantly decay when the interaction is gentle. Once higher harmonic amplitudes are accessible, one can also detect variations in higher harmonic phase shifts. In this work, variations in sample composition, or chemistry, here modelled through the Hamaker constant, have been shown to lead to variations in higher harmonic phase shifts and amplitudes. In particular, variations in the Hamaker constant of the order of 1020 J can induce higher harmonic phase shifts in the order of 10°. This is provided the higher harmonic amplitudes are small enough, i.e., about 1–10 pm. These small variations in phase shift would suffice to distinguish between metals such as gold, silver or copper [40]. Higher harmonic phase shifts also provide the means to decouple the true topography from an apparent topography, which is induced by compositional variations. Furthermore this outcome should still be valid in standard bimodal imaging. Overall, the proposed approach, and variations, might ultimately fulfil the promise of rapid chemical identification with multiple contrast channels while simultaneously exerting only gentle forces on samples. Still it has to be acknowledged that, experimentally, it is expected that technical issues might arise from the multiple excitation of exact frequencies and from the set-up required to detect variations in higher harmonic phase. In particular, the set-up would require the generation of exact harmonic external drives to bring the harmonic amplitudes above the noise level while keeping them small enough to provide enough phase contrast. This last point is relevant since it has been shown that higher harmonic amplitudes should remain in the sub-100-pm range for the higher harmonic phase shifts to be significantly large, i.e., above 0.2°, in response to variations in the tip–sample force. On the other hand, an analysis of thermal fluctuation that exploits the equipartition theorem has also indicated that thermal noise should be of the order of 0.1–1.0 pm close to the higher harmonics modes. The implication is that the working amplitudes should lie in the range of 1 to 100 pm. The noise analysis has also shown that there is an increase in sensitivity of the phase shift to the tip–sample force when frequencies are externally excited. Nevertheless, ultimately, only experimental practice, implementation, ingenuity and further theoretical advances in the field are to establish what the limits of this approach are.
Acknowledgements
This project was funded by Ministerio de Economía y Competitividad (MAT2012-38319). The artistic figure was designed by Maritsa Kissamitaki.
Stark, M.; Stark, R. W.; Heckl, W. M.; Guckenberger, R. Proc. Natl. Acad. Sci. U. S. A.2002,99, 8473–8478. doi:10.1073/pnas.122040599
Return to citation in text:
[1]
Sahin, O.; Quate, C.; Solgaard, O.; Atalar, A. Phys. Rev. B2004,69, 165416. doi:10.1103/PhysRevB.69.165416
Return to citation in text:
[1]
[2]
[3]
Sahin, O.; Quate, C.; Solgaard, O.; Giessibl, F. J. Higher-Harmonic Force Detection in Dynamic Force Microscopy. In Springer Handbook of Nanotechnology; Bhushan, B., Ed.; Springer Verlag: Berlin, Germany, 2010; pp 711–729. doi:10.1007/978-3-642-02525-9_25
Return to citation in text:
[1]
[2]
Payam, A. F.; Ramos, J. R.; Garcia, R. ACS Nano2012,6, 4663–4670. doi:10.1021/nn2048558
Return to citation in text:
[1]
[2]
[3]
[4]
Garcia, R.; Herruzo, E. T. Nat. Nanotechnol.2012,7, 217–226. doi:10.1038/nnano.2012.38
Return to citation in text:
[1]
[2]
[3]
[4]
Hutter, C.; Platz, D.; Tholén, E. A.; Hansson, T. H.; Haviland, D. B. Phys. Rev. Lett.2010,104, 050801–050804. doi:10.1103/PhysRevLett.104.050801
Return to citation in text:
[1]
Kawai, S.; Glatzel, T.; Koch, S.; Such, B.; Baratoff, A.; Meyer, E. Phys. Rev. Lett.2009,103, 220801–220804. doi:10.1103/PhysRevLett.103.220801
Return to citation in text:
[1]
[2]
[3]
[4]
Hembacher, S.; Giessibl, F. J.; Mannhart, J. Science2004,305, 380–383. doi:10.1126/science.1099730
Return to citation in text:
[1]
Basak, S.; Raman, A. Appl. Phys. Lett.2007,91, 064107–064109. doi:10.1063/1.2760175
Return to citation in text:
[1]
Xu, X.; Melcher, J.; Basak, S.; Reifenberger, R.; Raman, A. Phys. Rev. Lett.2009,102, 060801–060804. doi:10.1103/PhysRevLett.102.060801
Return to citation in text:
[1]
Sahin, O.; Magonov, S.; Su, C.; Quate, C. F.; Solgaard, O. Nat. Nanotechnol.2007,2, 507–514. doi:10.1038/nnano.2007.226
Return to citation in text:
[1]
Gadelrab, K.; Santos, S.; Font, J.; Chiesa, M. Nanoscale2013,5, 10776–10793. doi:10.1039/c3nr03338d
Return to citation in text:
[1]
[2]
[3]
Rodríguez, T. R.; García, R. Appl. Phys. Lett.2004,84, 449–551. doi:10.1063/1.1642273
Return to citation in text:
[1]
[2]
[3]
[4]
Stark, R. W.; Heckl, W. M. Rev. Sci. Instrum.2003,74, 5111–5114. doi:10.1063/1.1626008
Return to citation in text:
[1]
Preiner, J.; Tang, J.; Pastushenko, V.; Hinterdorfer, P. Phys. Rev. Lett.2007,99, 046102–046105. doi:10.1103/PhysRevLett.99.046102
Return to citation in text:
[1]
Dulebo, A.; Preiner, J.; Kienberger, F.; Kada, G.; Rankl, C.; Chtcheglova, L.; Lamprecht, C.; Kaftan, D.; Hinterdorfer, P. Ultramicroscopy2009,109, 1056–1060. doi:10.1016/j.ultramic.2009.03.020
Return to citation in text:
[1]
Turner, R. D.; Kirkham, J.; Devine, D.; Thomson, N. H. Appl. Phys. Lett.2009,94, 043901. doi:10.1063/1.3073825
Return to citation in text:
[1]
Raman, A.; Trigueros, S.; Cartagena, A.; Stevenson, A. P. Z.; Susilo, M.; Nauman, E.; Contera, S. A. Nat. Nanotechnol.2011,6, 809–814. doi:10.1038/nnano.2011.186
Return to citation in text:
[1]
Ramos, J. R. Desarrollos y applicaciones de la microscopía de fuerzas para el estudio de proteínas y de células cancerosas. Ph.D. Thesis, Universidad Autónoma de Madrid, Madrid, 2013.
Return to citation in text:
[1]
Reference 31
31.
Kiracofe, D.; Raman, A.; Yablon, D. Beilstein J. Nanotechnol.2013,4, 385–393. doi:10.3762/bjnano.4.45
Sahin, O.; Quate, C.; Solgaard, O.; Giessibl, F. J. Higher-Harmonic Force Detection in Dynamic Force Microscopy. In Springer Handbook of Nanotechnology; Bhushan, B., Ed.; Springer Verlag: Berlin, Germany, 2010; pp 711–729. doi:10.1007/978-3-642-02525-9_25
Sahin, O.; Quate, C.; Solgaard, O.; Giessibl, F. J. Higher-Harmonic Force Detection in Dynamic Force Microscopy. In Springer Handbook of Nanotechnology; Bhushan, B., Ed.; Springer Verlag: Berlin, Germany, 2010; pp 711–729. doi:10.1007/978-3-642-02525-9_25
Raman, A.; Trigueros, S.; Cartagena, A.; Stevenson, A. P. Z.; Susilo, M.; Nauman, E.; Contera, S. A. Nat. Nanotechnol.2011,6, 809–814. doi:10.1038/nnano.2011.186
Ramos, J. R. Desarrollos y applicaciones de la microscopía de fuerzas para el estudio de proteínas y de células cancerosas. Ph.D. Thesis, Universidad Autónoma de Madrid, Madrid, 2013.